美國戴安Dionex BorateTrap Column 047078 生物色譜柱
如果您對該產品感興趣的話,可以
產品名稱: 美國戴安Dionex BorateTrap Column 047078 生物色譜柱
產品型號: 047078
產品展商: 戴安Dionex
產品文檔: 無相關文檔
簡單介紹
戴安公司生產的儀器主要有離子色譜儀、高效液相色譜儀、快速溶劑萃取儀和氨基酸直接分析儀。其中離子色譜儀憑借著雄厚的技術優勢和方法已廣泛地應用于各個領域,其市場占有率達80%。全球與離子色譜相關的文章有90%均出自戴安公司生產的離子色譜儀。
戴安公司的快速溶劑萃取儀15分鐘之內僅用15毫升溶劑既可完成對10g樣品的萃取,該技術已在環境、**、食品和聚合物工業等領域得到廣泛應用。
美國戴安Dionex BorateTrap Column 047078 生物色譜柱
的詳細介紹
BorateTrap Column
The BorateTrap? is an in-line trap column, and is recommended for optimal performance during carbohydrate analysis.
-
In-line trap column removes borate contamination for HPAE-PAD.
-
Eliminates peak tailing for mannose, fructose, and sugar alcohols resulting from borate contamination.
Borate
contamination in eluents can cause a significant loss of peak
efficiency, especially for mannose, fructose, and reduced
monosaccharides. If borate is present in the eluent, it binds both to
the anion-exchange column and to the carbohydrate analytes. The
carbohydrate-borate complex is less efficiently eluted from the
anion-exchanger than is the carbohydrate, resulting in peak tailing,
particularly where vicinal cis hydroxyl groups are present, such as for
mannose and sugar alcohols. The BorateTrap eliminates peak tailing for
mannose, fructose, and reduced monosaccharides, resulting from
contamination of the eluent from borate. Therefore, the BorateTrap is
highly recommended for optimal performance during carbohydrate analysis.
The
BorateTrap is a 4 × 50 mm column placed between the eluent pump and the
injection valve to remove borate contamination from eluents used for
HPAE-PAD. Because it is placed before the injection valve, it has no
effect on the efficiencies or retention times of the carbohydrate
analytes. This column is packed with a 20-μm high-capacity resin that
has high selectivity for borate. The resin used in the BorateTrap is
functionalized with a polyol capable of complexing borate in the
presence of hydroxide. Borate is a known contaminant in laboratory water
supplies. In chromatography, borate contamination of HPLC eluents may
be a result of degrading deionized water systems or as leachate from
borosilicate glassware.
Ordering Information
|
Borate Trap Column
|
047078
|